ChemNet
 
Previous article Next article Contents  

I. P. Andreeva, E. A. Yakovleva, V. G. Grigorenko, A. P. Osipov2

Peculiarities of the development of a multiparametric immunochromatographic test system based on the analysis of cardiomarkers

Abstract

An immunochromatographic test system was developed for the simultaneous analysis of two cardiomarkers: heart type fatty acid binding protein (h-FABP) and troponin I (TnI) in serum and whole blood. The main stages of creating a multiparameter test system were reported: the effect of the position of antibodies and the size of gold nanoparticles, influencing the results of the analysis. The limits of detection of the method were 10 ng/ml for h-FABP and 1 ng/ml for TnI; the coefficient of variation of the results of the analysis did not exceed 10%. Using the developed test system, the diagnosis of acute myocardial infarction (N = 9) was confirmed. The analysis time was 15 minutes.
Key words: immunochromatographic assay; human heart fatty acid binding protein; human troponin-I; acute myocardila infarction.
Moscow University Chemistry Bulletin.
2019, Vol. 60, No. 6, P. 417
   